System Overview
Instructions for Use Fabius MRI SW 3.n 23
Note that the selector lever is shown in the center
position. This ensures that both vaporizers are in
the locked position. Also, this is the recommended
position for the selector lever when moving the
Fabius MRI.
Dräger Auto Exclusion 2-Vaporizer Mount
(Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When one of the two vaporizers is selected for use
(opened), the interlock mechanism within that
vaporizer’s mounting system is activated automati-
cally, preventing the other vaporizer from being
used.
NOTE
Only Dräger vaporizers labeled as
“AUTO EXCLUSION” vaporizers are compatible
with the Dräger Auto Exclusion 2-Vaporizer Mount.