Operation
114
Instructions for Use Fabius MRI SW 3.n
Preparing for Storage or Transport
WARNING
When moving the anesthesia work-
station, remove all monitors and
equipment from the top shelf and
hinged arms, remove the absorber
system, vaporizers, and reserve gas
cylinders, push in the writing tray.
The anesthesia workstation should only be
moved by people who are physically capable
of handling its weight. Dräger recommends
that two people move the anesthesia worksta-
tion to aid in maneuverability.
Exercise special care so that the machine
does not tip when moving up or down incli-
nes, around corners, and across thresholds
(for example, in door frames and elevators).
Do not attempt to pull the machine over any
hoses, cords, or other obstacles on the floor.
WARNING
All the procedures described in this manual
require that any use of ferromagnetic tools
can only be done while the Fabius MRI is out
of the scanner room.
WARNING
Do not bring any ferromagnetic tools or equip-
ment into the scanning room.
Ferromagnetic objects (made of steel, iron, or
stainless steel) are strongly attracted to the
magnet and can become harmful projectiles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals