Assembly
54
Instructions for Use Fabius MRI SW 3.n
Installing the Drägersorb CLIC
Adapter*
The disposable absorber:
CLIC Absorber 800+,or
CLIC Absorber Free, or
Infinity ID CLIC Absorber 800+, or
Infinity ID CLIC Absorber Free
can be used on the Fabius MRI by using the
Drägersorb CLIC adapter.
For information on installing the Drägersorb CLIC
adapter, consult its Instructions for Use.
To click the absorber into place:
1 Press the button; the mounting swings open.
z Before fitting, shake the disposable absorber,
e.g. by turning it upside down several times in
order to loosen up the soda lime.
z Remove seal from new disposable absorber.
z Slide the new disposable absorber onto the
mounting and
2 Push the disposable absorber in the direction of
the machine until it engages.
*optional
WARNING
The disposable absorber must be clicked into
place before switching on the Fabius MRI, so
that the absorber is included in the leak and
compliance test for the machine.
Otherwise breathing circuit leaks could be
undetected.
076
1
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals