Instructions for Use Fabius MRI SW 3.n 155
Configuration
1 During the calibration period, the word “CAL”
replaces the O
2 value in the oxygen monitoring
window. Calibration time is approximately
15 seconds. Upon successful completion of the
calibration, the O
2 measurement value will be
restored.
If the O
2 sensor cannot be calibrated, replace the
O
2 capsule in the O2 sensor housing (see
page 189).
If the O
2 sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
Des Comp On/Off
CAUTION
The Des Comp On/Off function is not relevant for
MRI use.
CAUTION
The Des Comp On/Off parameter remains in the
Off position on the Fabius MRI.
WARNING
The Desflurane vaporizer is not useable in an
MR environment. In an MR environment func-
tionality of the Desflurane vaporizer will be
compromised.