Instructions for Use Fabius MRI SW 3.n 155
Configuration
1 During the calibration period, the word “CAL
replaces the O
2 value in the oxygen monitoring
window. Calibration time is approximately
15 seconds. Upon successful completion of the
calibration, the O
2 measurement value will be
restored.
If the O
2 sensor cannot be calibrated, replace the
O
2 capsule in the O2 sensor housing (see
page 189).
If the O
2 sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
Des Comp On/Off
CAUTION
The Des Comp On/Off function is not relevant for
MRI use.
CAUTION
The Des Comp On/Off parameter remains in the
Off position on the Fabius MRI.
WARNING
The Desflurane vaporizer is not useable in an
MR environment. In an MR environment func-
tionality of the Desflurane vaporizer will be
compromised.
515
1
503

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals