EasyManuals Logo

Dräger Fabius User Manual

Dräger Fabius
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
Configuration
134
Instructions for Use Fabius MRI SW 3.n
Calibrate Flow Sensor
z Press the »Calibrate Flow Sensor« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
z Follow the instructions on the screen:
Close all fresh gas control valves.
Remove expiratory hose from breathing sys-
tem.
To start Flow Sensor Calibration, press rotary
knob.
z When the calibration begins, the instructions
are removed, and a »Flow Calibration in
progress« message is displayed above the
Standby soft keys.
z When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »Flow Calibration completed - recon-
nect expiratory hose« or »Flow Calibration
Failed«.
Troubleshooting Flow Calibration Failure
If the flow sensor cannot be calibrated, retry the cal-
ibration.
If the flow sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
599
525
526
527
528

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals