Instructions for Use Fabius MRI SW 3.n 137
Configuration
Compliance Test Results
The system compliance test determines the current
compliance of the patient system with any filters,
hoses, and Y-piece. The compliance value is used
during volume controlled ventilation for applying a
tidal volume matching accurately the set tidal vol-
ume. A system compliance value of up
6.5 mL/cmH
2O (6.5 mL/hPa) is posted as
»PASSED« on the leak test result screen, and the
compliance value is also displayed on the Standby
screen.
Ventilator Leak Test Results
If a ventilator leak test is posted as »PASSED«, the
value is displayed on the leak test result screen and
the Standby screen.
System Leak Test Results
System leak test results are posted on the leak test
result screen.
Measured ventilator
leak [mL/min]
Displayed result
[mL/min]
150 measured value and
PASSED
151 to 250 measured value and
FAILED
>250 >250 and FAILED
Measured system leak
[mL/min]
Displayed result
[mL/min]
250 measured value and
PASSED
251 to 350 measured value and
FAILED
>350 >350 and FAILED
116
O2+
Ventilator
Fresh-Gas
Decoupling
Vaporizer
Flow Tube
Oxygen
Flush
Sample Line: closed
Absorber
APL
Bypass
APL
Valve
Setting
PMAX/
PEEP
Bag
Scavenging
Y-piece: closed
115
O2+
Ventilator
Fresh-Gas
Decoupling
Vaporizer
Flow Tube
Oxygen
Flush
Sample Line: closed
Absorber
APL
Bypass
APL
Valve
Setting
PMAX/
PEEP
Bag
Scavenging
Y-piece: closed