Instructions for Use Fabius MRI SW 3.n 161
Fault-Cause-Remedy
z Adjust the APL pressure limit for the desired
inspiratory plateau pressure.
z Press the O2 flush button on the Fabius MRI as
required to sufficiently inflate the breathing bag.
z Manually ventilate the patient by squeezing the
breathing bag.
z Contact DrägerService or your local authorized
service organization before using the ventilator.
NOTE
After toggling the main power switch, the
Fabius MRI will perform its diagnostic tests. During
the diagnostic tests, manual ventilation is possible.
If the diagnostic tests result in FUNCTIONAL, the
Fabius MRI will automatically switch to ManSpont
mode if a fresh-gas flow is detected. Fabius MRI
respiratory monitoring is available. If the diagnostic
tests result in NON-FUNCTIONAL, manual venti-
lation is still possible but Fabius MRI respiratory
monitoring is not available.
NOTE
In ventilator override situation, the ventilator piston
assembly position may not be locked, as in
ManSpont mode. As a result, airway pressure may
initially push the piston back to its limit stop, thus
increasing the volume of the breathing circuit. It
may be necessary to press the O
2 flush button
again to reinflate the breathing bag.