Instructions for Use Fabius MRI SW 3.n 197
Technical Data
Gas supply from supplementary O
2 and N2O cylinders (with pin-index connections)
Cylinder Connections Pin-indexed hanger yokes (CGA V-1-1994)
Cylinder Gas Pressure
(typical full loads at 70 ºF, 21 ºC)
O
2, AIR 1900 psi (131 kPa x 100)
N2O 745 psi (51.3 kPa x 100)
Cylinder Gauges Conform to ASME B40.1 Grade B
Cylinder Gauge Range O
2 0 to 3000 psi (206.8 kPa x 100)
N2O 0 to 3000 psi (206.8 kPa x 100)
AIR 0 to 3000 psi (206.8 kPa x 100)
Compressed gas supply at workstation inlet
Dew point >5 °C (41 °F) at ambient temperature
Oil content <0.1 mg/m
3
Particles dust-free air (filtered with pores <1 µm)
Internal Regulator Safety Relief Valve Pressure 70 psi (4.8 kPa x 100) opening pressure
Protection Class I, in accordance with IEC 60601-1
Applied parts:
Breathing system (nozzles, breathing hoses) Type BF
Ingress of Fluids IPX0
Power supply, Rating non-configurable
100 to 240 VAC, 50/60 Hz, 8.7 A including auxiliary
power sockets
Rechargeable batteries
Rating: 24 V; 3.5 Ah
Type: sealed, gelled lead-acid
Recharging time: 16 hours on the mains or full operation time
Operation time with fully charged batteries: 45 minutes, minimum

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min