Instructions for Use Fabius MRI SW 3.n 73
Assembly
WARNING
For operation in magnetic fields, it is not per-
mitted to connect the vaporizer via hose con-
nectors or tapered connectors with the
anesthesia workstation.
WARNING
Risk of injury
Do not use ferromagnetic keyed filler or drain
adapters or tools when the filling or draining
procedure is carried out in magnetic fields.
Ferromagnetic adapters or tools can be
moved by magnetic attraction.
WARNING
Risk of injury
Vaporizers have to be mounted and removed
outside of the MR room.
Do not exchange any vaporizer in the MR
room.
Vaporizers can be moved by magnetic attrac-
tion.
WARNING
Risk of injury
Do not use filling adapters without a label
“MRI”. Adapters can be moved by magnetic
attraction.
CAUTION
Only Vapor 2000 vaporizers can by used on the
Fabius MRI in MRT scanner rooms.
CAUTION
Risk of injury
Dräger metal keyed filler adapters and Dräger fill
adapters are labeled with “MRI” are not ferromag-
netic.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals