DANGER!
To avoid the risk of the system tipping over
, the system must be firmly
connected to a suitable wall or floor.
NOTICE!
Availability and use of protective equipment
–
Protective gear is not part of the scope of delivery.
– The operator must provide safety glasses and gloves.
These items must be stored in a suitable location.
CAUTION!
– Any connection and repair work on the unit must be undertaken by
authorised experts.
– Before starting any work on electrical parts, switch of
f the power supply to
the controller and the signal conducting machines.
– Suitable protective clothing must be worn during maintenance and repair
work.
– These safety requirements and hazardous materials regulations for working
with chemicals and, in particular, the instructions pertaining to the
respective safety data sheets must always be followed.
The scope of delivery does NOT include any system separation
for safeguarding in accordance with the drinking water requirement
DIN EN 1717
!
See
Ä
Chapter 11.2 ‘Optional equipment ULTRAX Compact’ on page 77
It is essential that you observe the standards and regulations that apply
to you!
If required, please contact us (
Ä
Chapter 1.10 ‘Contact’ on page 13 ).
7.2 Installation prerequisites
1. Ensure there is sufficient space for wall mounting
2. On a dry-wall structure, use the proper-sized anchor mount (a hollow-wall type
anchor mount) to attach the device.
3. Make sure the mains power supply is working and available.
4. Provide a compressed air connection (Ø ½", 0.6 MPa / 6 bar, oil-free).
5. Pay attention to the fresh-water connection requirements.
Prevent equipment faults by fitting a mud guard upstream.
Installation
53 MAN046590 Rev. 6-03.2022