12.2.3  Range selection
When the camera is turned on, the temperature range is automatically 
selected, but may  have to be adjusted. When the environmental condi-
tions change, the range selection may also need to be changed.
NOTE: The range selection shall reect the temperature of the scene (not the place where 
the camera is positioned).
Selecting temperature range
Step Action
1 Select Setup/IR menu to open the IR Camera menu system.
2 Select Image/ Range….
3 Use the LEFT/RIGHT navigation buttons to select the correct temperature range.
4 Press the ENTER button to conrm.
5 Press the CANCEL button to exit the IR Camera menu system.
12.2.4  NUC
The NUC Non-Uniformity Correction) function performs an internal cali-
bration of the camera.
NUC can be performed in three different ways. With Internal NUC, the 
image is calibrated against a shutter inside the camera (Internal shutter). 
With External NUC, either the lens cover External shutter) or the scene 
(Shutter off) is used as thermal reference.
The normal procedure is to start with an Internal NUC. If an acceptable 
image quality is not achieved, an External NUC against the lens cover or 
against the scene may improve the image.
NOTE: When the camera is started, a NUC is needed every 5 minutes. After about 30 
minutes, a NUC is only needed when the quality of the image is degenerated.