Ranger HRC™ operator´s manual – Optimizing the IR Image
164 Publ. No. TM 614 006 699 Rev B – ENGLISH (EN) – Oct 30. 2008
12.1.3 Range selection
When the IR Camera is turned on, the temperature range is automatically
selected, but may have to be adjusted. When the environmental condi-
tions change, the range selection may also need to be changed.
NOTE: The range selection shall reect the temperature of the scene (not the place
where the camera is positioned).
Selecting temperature range
Step Action
1 Press the ENTER button to enter MENU mode.
2 Select IR menu from the Setup menu and press the ENTER button. The menu system
of the IR Camera is opened.
3 Select Image and press the ENTER button.
4 Select Range… and press the ENTER button. The Range dialog box is displayed.
5 Use the LEFT/RIGHT navigation buttons to select the correct temperature range.
6 Press the ENTER button to conrm.
7 Press the CANCEL button to exit the IR Camera menu system.
12.1.4 NUC
The NUC Non-Uniformity Correction) function performs an internal ca-
libration of the camera.
NUC can be performed in three different ways. With Internal NUC, the
image is calibrated against a shutter inside the camera. With External
NUC, either the lens cover or the scene is used as thermal reference.
An Internal NUC normally gives a good image quality.
NOTE: When the camera is started, a NUC is needed every 5 minutes. After about 30
minutes, a NUC is only needed when the quality of the image is degenerated.

Table of Contents