160
_____________________________________________________________________________________________
10.1 RS-232C Interface
______________________________________________________________________________________________
NOTE
This unit starts analysis after confirming the delimiter.
(2) Response format
When a command is sent to 3153 processes the command. When processing
is completed, 3153 always returns a response.
When there is no information from 3153
No response
When there is information from 3153 (measurement values, etc.)
Response character string related to the command ( +delimiter )
(3) Parameters
The 3153 uses parameters composed of decimal numbers. There are three
different formats for decimal data: NR1, NR2, and NR3. Each has two
values: one with a code and one without it. A value without a code is
regarded as a positive number. If a number exceeds the accuracy resolution
of the 3153, the value is rounded up or down.
NR1 format: Integer data
+12, -23, 34
NR2 format: Fixed point numbers
+1.23, -23.45, 3.456
NR3: Floating point number
+1.2E3, -2.3E-4, 3.4E5
The term "NRf format" includes all these three formats.
Each 3153 command designates a format.
(4) Delimiter
Depending on transmission direction, the delimiter is as follows.
From computer to 3153: CR or CR + LF
From 3153 to computer: CR + LF