_____________________________________________________________________________________________
10.8 GP-IB Command Reference
______________________________________________________________________________________________
10.8.1 Command Reference
Syntax
ESE <data>
<data>
0to255
(NR1)
Function
Sets the standard event status enable register (SESER) to a bit pattern which is used to mask
the standard event status register (SESR).
The numerical value can be in NRf format, but any digits after the decimal point will be
rounded.
When the power is turned on, the data is reinitialized to zero.
Example
Bits 2 and 4 of SESER are set to 1.
Transmission ESE 20
Error
If <data> is other than numerical value described above, an execution error occurs.
Standard event status enable re
g
ister
(
SESER
)
128 64 32 16 8 4 2 1
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
PON URQ CME EXE DDE QYE RQC OPC
10.8 GP-IB Command Reference
ESE
Sets the standard event status enable register.
The commands specified below can be used through interface function.
The commands are the same as those used for RS-232C. See Section 10.3.1
Common Command. Those specified below are the commands that can be
used for GP-IB only.