EasyManua.ls Logo

Hioki 3153 - Setting the Upper (Lower) Limit Value

Hioki 3153
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
_____________________________________________________________________________________________
3.3 "SETTING" State
______________________________________________________________________________________________
3.3.2 Setting the Upper (Lower) Limit Value
Upper (Lower) Limit Value: 20 mA
Upper-limit value
Lower-limit value: 1.0 mA
Lower-limit value
Lower-limit value: OFF
Lower-limit value
NOTE
The setting resolution of the upper (lower)-limit value is 0.1 mA at 0.1 mA to 9.9
mA, and 1 mA at 10 mA to 100 mA.
The current measurement resolution during a test depends on the set upper-limit
value: 0.01 mA at 0.10 mA to 10.00 mA and 0.1 mA at 10.1 mA to 100.0 mA.
If the set lower-limit value is greater than the upper-limit value, the test cannot be
started, even if the
START
key is pressed. In such a case, correct the upper- or
lower-limit value.
When the AC withstand-voltage mode is changed to the DC withstand-voltage
mode, if the set upper-limit value is 10 mA or greater and the set lower-limit
value is 9.9 mA or greater, the values for DC withstand-voltage testing will
automatically be changed to 10 mA for the upper-limit, and 9.9 mA for the
lower-limit.
1. If no flashing cursor is displayed in the withstand-voltage mode READY
state, press either the / key to display the cursor in the upper (lower)-
limit value position.
2. Set the upper (lower)-limit value using the
/
keys.
The upper (lower)-limit value will change by 0.1 mA (by 1 mA at 10 mA
to 100 mA). While holding down the
SHIFT
key, press the
/
keys.
(by10mAat10mAto100mA).
The following is the setting range:
The upper-limit value can be set between 0.1 and 100 mA. The lower-
limit value can be set between 0.1 and 10 mA. (DC withstand-voltage)
The lower-limit value can be set between 0.1 and 99 mA. The lower-limit
value can be set between 0.1 and 9.9 mA. (DC withstand-voltage)
If no lower-limit value is required, turn off the
ON/OFF
key. Upper-limit
value can not be turn off.
3. To finalize the test settings, press the
STOP
key. The settings are
confirmed and the unit returns to the READY state.

Table of Contents

Related product manuals