5.3 "SETTING" State
______________________________________________________________________________________________
Forcibly terminate
Make insulation-resistance mode
test settings
Make withstand-voltage mode
W
→
I
/
I
→
W
key
NOTE
If the test time is set to OFF in the withstand-voltage mode or insulation-
resistance mode, the test mode with its time set to
OFF
is displayed (
OFF
flashes)
upon startup of the test in the Auto test mode, and the test cannot be started.
Return to withstand-voltage or insulation-resistance mode, and set the test time to
ON
.
If the set lower-limit value ≧ upper-limit value, an item of comparative value
flashes upon startup of the test, and the test cannot be started.
In the insulation-resistance mode, if the set delay time ≧ test time, time-setting
items flash upon startup of the test, and the test cannot be started.
・Note that the unit does not start test if the test-voltage value is greater than the
out
ut-volta
e restrictin
value. See Section 7.9.
Flashing
5.3 "SETTING" State
In Auto test mode, test settings cannot be changed. To change test settings,
the withstand-voltage and insulation-resistance modes must be in the
SETTING state. (See Section 3.3 and 4.3)
After you have made the various test settings, select Auto test mode.
If you try to start a test in the Auto test mode, the test mode when the test
time is switched
OFF
is displayed (
OFF
flashes), and the test cannot be
started.