232
_____________________________________________________________________________________________
12.5 Removing the Lithium Battery
______________________________________________________________________________________________
WARNING
To avoid electrocution, turn off the power switch anddisconnect the
power cord and measurement cables before removing the lithium
battery. When disposing of this product, remove the lithium battery and
dispose of battery and product in accordance with local regulations.
CALIFORNIA, USA ONLY
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
12.5 Removing the Lithium Battery
1. Turn OFF the power switch on the unit, and disconnect the power cord
and probe.
2. Remove the 2 screws from the bottom of the unit using a Phillips
screwdriver.
3. Remove the 4 legs (fastened with 4 screws) from the back of the unit
using a Phillips screwdriver.
4. Remove the cover by sliding it backward.
5. The battery holder is in the position illustrated in the figure. Insert a
pointed object such as tweezers between the battery and the battery holder,
and remove the battery by lifting the battery.