30
_____________________________________________________________________________________________
3.3 "SETTING" State
______________________________________________________________________________________________
3.3.6 Examples of Settings
Test-voltage value: 2.00 kV
Upper-limit value: 20 mA
Lower-limit value: OFF
Test time: 60.0 s
Ramp-up time: 10 s
Ramp-down time: OFF
Test-voltage type: AC 60 Hz
Values currently set
Values to be set
Test-voltage value 1.50 kV Test-voltage value 2.00 kV
Upper-limit value 100 mA Upper-limit value 20 mA
Lower-limit value 40 mA Lower-limit value OFF
Test time 120 s Test time 60.0 s
Ramp-up time 30 s Ramp-up time 10 s
Ramp-down time 20 s Ramp-down time OFF
Test voltage type AC 50 Hz Test-voltage type AC 60 Hz
This section describes the procedure for making withstand-voltage test
settings under the following conditions.
The 3153 is in the READY state in withstand-voltage mode.
(1) Changing to the SETTING state
Press the / keys to switch to the SETTING state. The
READY
light will go
out, and the flashing cursor will be displayed at the position where the test-
voltage value is indicated.