EasyManuals Logo

Keysight N9038A Service Guide

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
140 Keysight N9038A MXE Service Guide
Input Selection & Level Control
Input Selection & Level Control Section Troubleshooting
12.Verify that the signal level on the spectrum analyzer is at 20 dBm
(±0.25 dB), as shown in Figure 4-6, allowing for any additional test cable
loss.
If the signal level is incorrect the most likely cause of the problem is the A9
Attenuator A assembly itself. However, before replacing an A9 Attenuator
A assembly for any failure, refer to Chapter 8,Front End Control. for
information on how to verify the control signals to it.
Figure 4-6 A9 Attenuator A AC/DC Input Coupling
13.Turn on the Marker Delta function of the spectrum analyzer being used.
14.Change the instrument input coupling to AC by pressing Input / Output,
RF Input, RF Coupling AC.
15.Verify that the signal level on the spectrum analyzer drops by at least 5 dB
from where it was when the RF Coupling was set to DC, as shown in Figure
4-6.
If the signal level is incorrect the most likely cause of the problem is the A9
Attenuator A assembly itself. However, before replacing an A9 Attenuator
A assembly for any failure, refer to Chapter 8,Front End Control. for
information on how to verify the control signals to it.

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals