EasyManuals Logo

Keysight N9038A Service Guide

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #141 background imageLoading...
Page #141 background image
Keysight N9038A MXE Service Guide 141
Input Selection & Level Control
Input Selection & Level Control Section Troubleshooting
RF Input Selection
16.Move the signal generator to the A9 Attenuator A RF Input #1. Refer to
Figure 4-4 or Figure 4-5.
17.Tune the signal generator to 1 GHz with an amplitude of 20 dBm.
18.Tune the spectrum analyzer to 1 GHz with a span of 1 MHz.
19.Verify that the signal level on the spectrum analyzer is at 20 dBm
(±0.25 dB), as shown in Figure 4-7, allowing for any additional test cable
loss.
If the signal level is incorrect the most likely cause of the problem is the A9
Attenuator A assembly itself. However, before replacing an A9 Attenuator
A assembly for any failure, refer to Chapter 8,Front End Control. for
information on how to verify the control signals to it.
Figure 4-7 A9 Attenuator A RF Input #1
If there is a particular frequency of interest use that instead of the 1 GHz used in this example.

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals