Keysight N9038A MXE Service Guide 569
Assembly Replacement Procedures
RF Area - Option 544
Replacement
1. Attach the two ribbon cables to the switches. Torque the SW1 ribbon
cable screws to 9 inch-lbs.
2. Refer to Figure 18-33. Place the switch/bracket assembly onto the chassis
and use the three screws (1) to attach the bracket to the chassis. Torque
the screws to 9 inch-lbs.
3. Refer to Figure 18-32. Replace the semi-rigid cables in this order: W39,
W31, W36, W38, W27, W30, W28, W37, and W43. Torque all cables to
10 inch-lbs.
4. Refer to Figure 18-24. Position the Chassis Right Side Outer bracket onto
the chassis and replace the eleven screws (0515-0372). Torque to
9 inch-pounds.
5. Replace the top brace. Refer to the Top Brace and Power Supply Bracket
replacement procedure.
6. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
7. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.