8.1 Transducer
245
Chapter 8
CAUTION
•••••••••••••••••••••••••••••••••••••
Do not use a transducer whose acoustic lens or
housing is abnormally hot. Doing so may cause a
burn injury.
Do not use a hard brush for cleaning or disinfecting
the transducer. Doing so may damage it.
Do not immerse the transducer in a disinfection so-
lution longer than a specied time. If doing so, the
performance of the transducer could deteriorate.
Do not use non-specified chemicals (for disinfec-
tion or cleaning). Doing so may cause the perfor-
mance of transducers to decrease.
Turn OFF the system before attaching / detaching
or replacing a transducer.
When the transducer has to be laid down
temporarily during a study, place it on a stable and
level surface to prevent it from falling.
Do not connect anything other than transducers or
Three-Port Probe Unit designed for this system to
the transducer connector on the main unit.
When transporting a transducer, be sure to disin-
fect the transducer at the beginning, put it into a
plastic bag, and then put the transducer-enclosed
bag in a storage case. If do not doing so, the per-
formance of the transducer could deteriorate.
Use ultrasound gel that mainly consists of water.
Use of castor oil may deteriorate the transducer
and cause an electric shock.
Please select and use an appropriate ultrasound
gel on the market.
Before cleaning or disinfecting the transducer, re-
move it from this system. Otherwise there is a risk
of electric shock.
Wear protective gloves to prevent infection before
cleaning and disinfecting the transducer.
Be sure to comply with the usage and storage envi-
ronment conditions of the transducer.
If do not doing so, the performance of the trans-
ducer could deteriorate
.
Disinfect the transducer to prevent infection before
returning it to your service representative.
While the main unit is on stand by, be sure to
freeze the main unit.
If the live state is continued, there is a risk that the
transducer will overheat and cause burns.
Store the ultrasound gel as described in the in-
struction manual. Otherwise, there is a risk of infec-
tion.
When discarding transducers, disinfect or sterilize
the transducers, and then discard them as indus-
trial waste.
Be sure to follow the rules and regulations of the
local governing body, and request an approved
industrial waste disposal company to discard the
transducers.
•••••••••••••••••••••••••••••••••••••••••••••••••••••