304
11.7
Check List for Maintenance Check: Before and after use
Before and after use, inspect this system, according to the following check list.
Check the items below before turning on the system.
Name:
Date of Maintenance:
No. Item
Check o󰀨
(1)
The system has been put/kept under the specied environmental conditions.
(2)
Ambient temperature, relative humidity, and atmospheric pressure should meet the system require-
ments.
(3)
There should be no condensation.
(4)
There should be no deformation, damage, or dirt on the system.
(5)
There should be no loosening of screws.
(6)
There should be no damage or connector loosening in the power supply cord.
(7)
There should be no dirt and abnormality (projection, cracking, chasm, sharp edge or peeling) on the
surface of the transducer, cable, or connector.
(8)
There should be nothing blocking the air vent of the system and no obstacles near the transducer
cable, the wall mounted AC socket outlet and mains plug.
(9)
The system should be placed in a stable position during operation.
Check the items below after turning on the power.
No. Item
Check o󰀨
(1)
There must be no abnormal noise, abnormal smell, or heating.
(2)
No error message must be displayed.
(3)
There must be no intermittent display or dark areas on the ultrasound image.
(4)
The date and time must be indicated correctly.
(5)
The acoustic lens surface and housing of the transducer must not become abnormally hot. (Check
by touching the lens surface)
(6)
The touch panel must function normally.
(7)
The operation sound must not be overly loud.
Check the items below after turning o󰀨 the power.
No. Item Check o󰀨
(1)
There should be no deformation, damage, or dirt on the system.
(2)
There should be no loosening of screws.
(3)
There should be no damage or connector loosening in the power supply cord.
(4)
There should be no dirt and abnormality (projection, cracking, chasm, or peeling) on the surface of
the transducer, cable, or connector.
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Please refer to "7.2.2 Cleaning Procedures" before cleaning dirt on the system.
Please refer to "8.1.6 Cleaning and Disinfection" before washing o󰀨 residues on the transducer.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••