EasyManua.ls Logo

Konica Minolta SONIMAGE HS2 - 8.2.8 Selection of Biopsy Bracket

Konica Minolta SONIMAGE HS2
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
264
8.2 Biopsy
8.2.8 Selection of Biopsy Bracket
Here is the information about the biopsy bracket symbols and angles corresponding to the transducers to be applied.
No. Transducer number Biopsy bracket symbol
Biopsy needle insertion angle (deg)
Depth values in parentheses
θA θB θC
(1) C5-2 Multi
35.1
[50 mm]
23.1
[90 mm]
(2) L18-4 Multi
60.0
[15 mm]
47.4
[25 mm]
38.5
[35 mm]
(3) L14-4 Multi
54.6
[20 mm]
37.2
[40 mm]
(4) L11-3 Multi
61.1
[15 mm]
49.2
[25 mm]
37.1
[40 mm]
Example) in the (1) case Example) in the (2), (3) and (4) case
X
Y
θ°
Depth
X
Y
θ°
Depth

Table of Contents