240
7.3 Periodic Inspection Check
Konica Minolta provides the periodic maintenance inspection service for a minor evaluation fee.
Contact your service representative for details.
Inspection contents Inspection items Inspection cycle
1. Cleaning
Vent holes
Main unit
1 year (at periodic inspections)
2. Electrical safety check Protective earth resistance, etc. 1 year (at periodic inspection)
3. Mechanical safety check
Casters (when connected to the Pole Cart2), LCD
monitor, appearance of transducer
Exterior parts, holder
1 year (at periodic inspection)
4. Image check and record
Ultrasound image in each mode
Ultrasound image using the transducer
1 year (at periodic inspection)
5. Basic Maintenance
Self-diagnosis check (System Test, System Mainte-
nance and Hardware/Software diagnosis) in mainte-
nance mode.
1 year (at periodic inspection)
6. Consumable parts
Fan
40,000 hours - Check time using the maintenance
function.
1 year (at periodic inspection) - Check
necessity of replacement.
Battery
1 year
1 year (at periodic inspection) - Check
necessity of replacement.

Table of Contents