80
2.3 Power ON/OFF
z
Shutdown function
Pressing the power switch will display the message
asking you to select the desired shutdown type.
List of shutdown conrmation message buttons
Button Functions
[Standby] Puts the system into standby state.
[Shutdown] Shuts down the system.
[Cancel] Cancels the shutdown.
When the security function is enabled, the [Logout]
button is displayed on the shutdown conrmation mes-
sage.
To configure the security function, refer to "User Ac-
count Settings" of the "Operation Manual: Applica-
tions".
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• For longer and continued use of the system, it is
recommended to shut down the system at least
once every three days.
• If you use the system when the remaining battery
level is low, do not remove the AC adapter before
shutting down the system. Doing so may cause a
failure of the system.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2.3.3 Standby
z
Power switch > Shutdown conrmation
message > [Standby] button
• Executes Standby.
• Executes [End Study] instead during examination
and ends the examination.
• Saves the current system status in the memory, and
enters the energy-saving mode.
• To start up the system in a standby state, press
the power switch, which starts up the system in the
same manner as it does in the normal power-on
state.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• If the monitor is closed, the standby processing is
executed.
• Do not disconnect the battery during the standby
state or leave the system in the standby state for a
long period of time. If the battery runs out, memory
corruption can occur, and quick startup from a
standby condition becomes impossible.
•••••••••••••••••••••••••••••••••••••••••••••••••••••