301
Chapter 11
This system is classied with respect to safety characteristics.
(1) According to the type of protection against electric shock:
System: CLASS I EQUIPMENT
(2) According to the degree of protection against electric shock:
EQUIPMENT WITH TYPE BF APPLIED PARTS
(3) According to the degree of protection against harmful ingress of water:
Mainframe: IPX0 (enclosed EQUIPMENT without protection against ingress of water)
Transducer: IPX7
Refer to "8.1.3 Transducer Part Names".
Foot Switch: IPX8
(4) According to the degree of safety of application in the presence of FLAMMABLE
ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE:
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or
WITH OXYGEN OR NITROUS OXIDE
(5) According to the mode of operation:
CONTINUOUS OPERATION
11.5 Safety Classication