248
8.1 Transducer
z
Linear type (optional)
Name LINEAR PROBE L11-3
Scan system Electronic linear scanning
Biopsy function Available
Weight Approximately 520 g
Cable length Approximately 1950 mm
Operating environment
Ambient temperature 10 °C to 40 °C
Relative humidity 30 % to 80 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa
Shipment and storage conditions
Ambient temperature -20 °C to 60 °C
Relative humidity 20 % to 90 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa
z
Linear type (optional)
Name LINEAR PROBE HL18-4
Scan system Electronic linear scanning
Biopsy function Not available
Weight Approximately 420 g
Cable length Approximately 2200 mm
Operating environment
Ambient temperature 10 °C to 40 °C
Relative humidity 30 % to 80 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa
Shipment and storage conditions
Ambient temperature -20 °C to 60 °C
Relative humidity 20 % to 90 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa
z
Linear type (optional)
Name LINEAR PROBE WL13-3
Scan system Electronic linear scanning
Biopsy function Not available
Weight Approximately 520 g
Cable length Approximately 1950 mm
Operating environment
Ambient temperature 10 °C to 40 °C
Relative humidity 30 % to 80 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa
Shipment and storage conditions
Ambient temperature -20 °C to 60 °C
Relative humidity 20 % to 90 % (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa