11/8/00 AUD (Optional) 12-7
Use the the up/down arrows

to highlight “THD” and
press the check key
.
The THD display shows the results of the THD calculations.
THD information is only available
on screen at the time of the measure-
ment and is not stored with a file or
printed to a report.
Results are reported in dB down difference and units of per-
centile. The dB down difference is calculated by subtracting
the harmonic level from the fundamental level. The percent-
age is equal to one hundred times ten raised to the power of
the result of the dB down difference divided by twenty.
Entering the THD display will auto-
position the cursor to the fundamen-
tal frequency on the FFT screen.
THD is calculated by scanning the full spectrum to find the
fundamental frequency and selecting the highest peak. The
actual frequency is extrapolated from the fundamental bin
frequency and the highest sideband bin frequency. The sec-
ond harmonic is found by doubling the fundamental fre-
quency, then finding the nearest bin frequency and summing
its level with it’s side-band levels. The third harmonic is
found by tripling the fundamental frequency, then finding
the nearest bin frequency and summing its level with it’s
side-band levels. The fourth and higher harmonics are found
by the process described above for the second and third har-
monics.
THD+n is calculated by summing the energy of the funda-
mental bin with the five bins on each side, then subtracting
the energy sum of all the other levels.
The THD as calculated in the 824
satisfies all of the requirements for
ANSI S3.6.
For THD to be calculated, the following criteria must be
met:
The window must be in Hanning mode after an overall
reset. If not, “Available only in Hanning Window” is
shown.
THD+n percentile
Harmonic percentiles
4th & higher harmonic
percentile
4th & higher harmonic
Harmonic dB down
THD+n dB down
Fundamental frequency

Table of Contents

Other manuals for Larson Davis System 824

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Larson Davis System 824 and is the answer not in the manual?

Larson Davis System 824 Specifications

General IconGeneral
BrandLarson Davis
ModelSystem 824
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals