EasyManuals Logo

LeCroy SDA User Manual

LeCroy SDA
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
SDA Operator’s Manual
12. Under "3D settings," touch inside the Quality field and select an image quality from the pop-
up menu: wire frame, solid, or shaded.
13. For each axis, touch inside the data entry field and enter a value from -90° to +90°.
14. To turn off persistence for an individual channel, touch the left-most persistence mode button
. To turn off persistence for all channels, press the front panel
Analog Persist button
. This button toggles Analog Persistence on and off.
Screen Saver
The Windows screen saver is activated the same way as on a standard PC.
1. Minimize the instrument display by touching File in the menu bar, then Minimize in the drop-
down menu.
2. Touch Start down in the task bar.
3. Touch Settings in the pop-up menu.
4. Touch Control Panel.
5. Touch Display.
6. Touch the Screen Saver tab.
Moving Traces from Grid to Grid
You can move traces from grid to grid at the touch of a button.
Moving a Channel or Math Trace
1. Touch the descriptor label for the waveform that you want to move.
Example Descriptor Label (previous)
2. Now, touch the Next Grid button
.
Note: If you have more than one waveform displayed on only one grid, a second grid will open automatically when you
select Next Grid.
SDA-OM-E Rev H 101

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy SDA and is the answer not in the manual?

LeCroy SDA Specifications

General IconGeneral
BrandLeCroy
ModelSDA
CategoryMeasuring Instruments
LanguageEnglish