14 DICOM/HL7
Operator’s Manual 14 - 3
14.3 Worklist
When the DICOM basic package is configured and the Worklist server has been set, click
[Worklist] in the “Patient Info” screen to query or import the patient data.
The system supports: DICOM, HL7 and HIS.
Perform the following procedure:
1. In Patient screen, select [Worklist] in the bottom-left corner to enter the import screen.
2. Select the data source: select the server type (DICOM, HL7 or HIS), choose a Worklist server
in the “Server Type” drop-down list, and all the patient exam records are listed.
3. Enter the search condition:
Set query criteria, or select the search key, enter the keywords, then tap [Query] to search.
To reset the criteria, tap [Clear].
4. Select the desired patient record in the displayed patient list.
5. Select the desired patient and tap [Done]. The patient information is imported into the system
and then an exam is started.
14.4 MPPS
MPPS is used to send exam state information to the configured server. This facilitates the other
systems in obtaining the exam progress in time.
The status information is described as follows: