Operator’s Manual C - 1
C Trolley and Accessories
WARNING
DO NOT connect the trolley to the outlets with the same circuit breakers
and fuses that control the current to devices such as life-support systems. If
the trolley or the system malfunctions and generates over-current, or when
there is an instantaneous current at power ON, the circuit breakers and
fuses of the building’s supply circuit may be tripped.
Be sure to connect the equipotential wire before inserting the power plug
into the receptacle; be sure to remove the power plug from the receptacle
before disconnecting the equipotential wire; otherwise electric shock may
result.
When you connect another device to the trolley, you should use the
equipotential wire to connect each of equipotential terminals; otherwise
electric shock may result.
Connect the earth cable before powering ON. Disconnect the earth cable
after powering OFF. Otherwise, electric shock may result.
CAUTION
When moving the trolley with mounted system, please take care of the
connector of the power adapter in case of damage.
DO NOT insert fingers into the gap next to the plug in case of injury.
C.1 UMT-400 Trolley
C.1.1 Trolley Accessories
•Gel holder
Storage bin
Black/white video printer tray
Power cables
Towelette holster