7 - 2 Operator’s Manual
7 Visual Transient Elastography
7.2 Image Acquisition
Perform the following procedure:
1. Find the examination area.
Place the probe head to the right hypochondriac area located at the intersection of the
transverse line at the level of the xiphoid process and the midaxillary line, and keep the probe
button facing up.
2. Perform 2D real-time ultrasound examination on B mode or B + Color (Power) along the
intercostal space.
3. Adjust the position and size of ROI if necessary.
– Tap and hold the inner of ROI, and drag to adjust the position.
– Tap > [Image] > [ViTE], use [Fixed ROI] to adjust the size of ROI.
4. Select the number of automatic acquisition.