2 - 2 Operator’s Manual
2 System Overview
According to the degree of protection against harmful ingress of water:
The main unit is rated as IPX0
The probes are rated as IPX7
The foot switch (can be applied in the operating room) is rated as IPX8
The power adapter is rated as IPX1
According to the disinfection and sterilization method(s) recommended by manufacturer:
Equipment with disinfection and sterilization method(s) recommended by manufacturer.
According to the degree of safety of application in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation:
Continuous operation
According to the installation and use:
Portable equipment
Mobile equipment (when the system is installed on the mobile trolley)
Does the equipment has any defibrillation-proof applied parts:
Non-defibrillation-proof applied part
Permanently installed equipment or non-permanently installed equipment:
Non-permanently installed equipment
2.4 Product Specifications
NOTE:
The functions described in the operators manual may vary depending on the specific system
purchased.
2.4.1 Power supply
Voltage: 100 V - 240 V AC
Frequency: 50/60 Hz
Power input:
Main unit: 2.0 A
UMT-400 power input: 5.4-2.6 A
HMT-600 power input: 4.4-2.5 A
Battery voltage: 14.8 V
2.4.2 Environmental Conditions
Operating conditions
Ambient temperature: 0 °C to 40 °C
Relative humidity: 30% to 85% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Storage and transportation conditions
Ambient temperature: -20 °C to 55 °C