I - 6 Operator’s Manual
I Acoustic Output
I.8 Measurement Uncertainty
The total estimated measurement uncertainty (where the total uncertainty includes the uncertainties
in hydrophone response, measurement, calculation, and positioning) are:
I.9 References for Acoustic Power and Safety
“Bioeffects and Safety of Diagnostic Ultrasound” issued by AIUM in 1993
“Medical Ultrasound Safety” issued by AIUM in 1994
Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, June 27, 2019.
Center for Devices and Radiological Health.
Medical electrical equipment-Part 2-37: Particular requirements for the basic safety and
essential performance of ultrasonic medical diagnostic and monitoring equipment issued by
IEC in 2015
IEC 62359, Ultrasonics-Field characterization-Test methods for the determination of thermal
and mechanical indices related to medical diagnostic ultrasonic fields, 2017.
Acoustic Quantities Total Uncertainties (Standard)
Power 26.48% for non-scan modes
6.03% for scan modes
Frequency 0.22%
Pressure 13.01%
Ita 26.48% for non-scan modes
26.95% for scan modes
Ipa 26.5%
Mechanical Index 13.01%
Total Uncertainty for TIS Non-scan Modes: 26.48%
Scan-Modes: 6.03%
Total Uncertainty for TIB Non-scan Modes: 26.48% or 18.72%
Scan-Modes: 6.03%
Total Uncertainty for TIC Non-scan Modes: 26.48%
Scan-Modes: 6.03%