17 - 2 Operator’s Manual
17 System Maintenance
• Please clean the control panel periodically, otherwise the button may be blocked by dirt and the
system will be buzzing while the button makes no response.
Cleaning probes
Tools: mild soapy water , dry soft cloth, soft brush
Method:
1. Wipe out the dust attached to surface of probe head, connector and cable.
2. Use soft brush to brush the dust inside probe connector gently.
3. Remained stain or dust attached to surface of cable or surface of connector should be washed
out by cloth with a little soapy water, and then air-dry.
NOTE:
Don’t use cloth with water to clean the probe connector.
Cleaning holders
Tools: soft dry cloth, soapy water, soft brush.
Remaining stains should be wiped away using a cloth with clean or soapy water and the surface left
to air dry.
Perform the following procedure:
1. Use a soft dry cloth to wipe away dust attached to the inside, outside and gaps in the probe
holder. Use a soft brush to brush away dust or stains from the small intra-cavity probe holder or
its gap.
2. Remaining stains on the inside and outside of the holder should be wiped away using a cloth
with a little soapy water and then air dried.
Cleaning the display
NOTE:
Do not use chemical solvents, acid or alkaline solution to clean the display.
The ultrasound system has passed puncture test and leakage current test, and thus can be cleaned on
power-on status.
Tools: soft dry cloth, clean water or neutral detergent.
Clean the system in power off status; or in power on status, tap in the top-right corner of the
screen and select to lock the system for 10 seconds.
Clean the display surface with a soft dry cloth. Remaining stains should be wiped away using a
cloth with a little neutral detergent or clean water and then leave to air dry.
Cleaning the cover
Tools: soft dry cloth, soapy water
Use a soft dry cloth to clean the system’s cover. If the system is dirty, moisten the soft cloth with
neutral soapy water, wipe away any stains, and dry the system in the air.
NOTE:
Use a soft brush to brush away dust attached to all visible sockets or interfaces (such as probe
sockets, sockets or interfaces in the IO panel and power supply panel). Do not use a cloth and water.