J - 4 Operator’s Manual
J Acoustic Power Data and Surface Temperature Data
J.3 Acoustic Output Reporting Table for Track 3
J.3.1 C5-1s
Transducer Model: C5-1s
Imaging Mode: M-mode
Index label
MI TIS TIB TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value 1.31 0.14 0.24 0.23
Index component value
0.14 0.11 0.23 0.24
Acoustic
Parameters
p
r,α
at z
MI
(MPa)
1.73
P (mW) 11.17 11.17 11.17
P
1×1
(mW)
8.94 8.94
z
s
(cm)
1.86
z
b
(cm)
4.43
z
MI
(cm)
5.26
z
pii
,α
(cm)
5.26
f
awf
(MHz)
1.74 3.21 3.21 3.21
Other
Information
prr (Hz) 500.00
srr (Hz) /
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
168.07
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
95.57
I
spta
at z
pii
or z
sii
(mW/cm
2
)
186.04
p
r
at z
pii
(MPa)
2.32
Operating
control
conditions
Acoustic power 100% 100% 100% 100%
Display depth 22cm 22cm 22cm 22cm
Focus position 7.0cm 6.0cm 6.0cm 6.0cm
Working Frequency HRes Gen Gen Gen
PRF 500 1000 1000 1000

Table of Contents