J Acoustic Power Data and Surface Temperature Data
Operator’s Manual J - 13
Transducer Model: LFP5-1s
Imaging Mode: B-mode/Tissue Harmonic Imaging/Free Xros M/ViTE
Index label
MI TIS TIB TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value 1.28 0.09 0.57 0.57
Index component value
0.09 0.09 0.57 0.08
Acoustic
Parameters
p
r,α
at z
MI
(MPa)
2.14
P (mW) 81.20 86.83 86.83
P
1×1
(mW)
7.07 7.56
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.61
z
pii
,α
(cm)
1.61
f
awf
(MHz)
2.78 2.81 2.22 2.22
Other
Information
prr (Hz) 3,959.00
srr (Hz) 34.00
n
pps
0.00
I
pa,α
at z
pii
,α
(W/cm
2
)
169.60
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
10.48
I
spta
at z
pii
or z
sii
(mW/cm
2
)
14.49
p
r
at z
pii
(MPa)
2.44
Operating
control
conditions
Acoustic power 100% 100% 100% 100%
Display depth 16cm 16cm 16cm 16cm
Focus position 2.0cm 3.0cm 3.0cm 3.0cm
Working Frequency Gen Gen HGen HGen
PRF 3959 3959 4019 4019