V Series Operating Instructions 6 - 7
Patient Management Patient Info Dialog
6.2.6 Clearing Patient History
To clear a patient’s history without discharging them from the monitor:
1. Select the Header Bar (located above the top waveform).
The Patient Info dialog displays.
2. Select the Transfer/Discharge tab, then select the Clear History button.
3. A confirmation dialog displays the message Are you sure you want to clear the history data
of this patient?
4. Select No or Yes.
6.2.7 Placing a Patient in Standby
NOTE: When connected to a central monitoring system (i.e., central station),
and a patient is moved or removed from standby at the V 12/V 21
bedside monitor, the patient is moved or removed from standby at the
central station. Refer to the central station operating instructions for
additional information.
To enter the Standby mode:
1. Select the Header Bar (located above the top waveform).
The Patient Info dialog displays.
2. Select the Transfer/Discharge tab, then select the Standby button.
Placing a patient in the Standby mode does the following:
Temporarily suspends patient monitoring
Places alarms in permanent suspension
Places interval NIBP measurements in Pause mode
Shuts off the CO
2
pump
Removes all patient data from the screen and displays Standby mode on the display
6.2.8 Exiting the Standby Mode
To exit the Standby mode, select the touch screen. When in the Standby mode, the message
Touch screen to resume monitoring displays.
When exiting from the Standby mode to the monitoring mode, the monitor responds as follows:
Restores parameter measurement as previously configured.
Arrhythmia automatically relearns.
NIBP measurement does not automatically start. Restart NIBP measurements via the Start
button.
Restores normal data storage.
•The CO
2
pump automatically reactivates provided the sensor is in place.
All the alarms are activated.
The display settings are maintained during the Standby mode.
The monitor notifies the central station and other monitors when back in monitoring mode.