EasyManua.ls Logo

Mindray V Series - Power Specifications

Mindray V Series
587 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V Series Operating Instructions 29 - 53
Product Specifications Power Specifications
29.17 Power Specifications
Power Supply Type
AC power, battery
Whole-system consumption: maximum of 215 W.
AC Mains Power Source
AC power: 100~240 Vac, +/-10%
Current: 2.4 - 1.0 Amps
AC input power frequency: 60 Hz/50 Hz
Internal User-Removable Battery
Li-Ion battery (0146-00-0091-01, Mindray)
Number of batteries: 3 (maximum)
Li-Ion Battery Index
Voltage: 11.1VDC
Rated Battery capacity: 4.8 Ah (one battery)
Battery run time: >240min (three batteries)
Operating conditions: 25ºC for new fully-charged battery. Equipment configuration: 5-lead ECG,
Impedance Respiration, Masimo SpO
2
, Temperature, IBP1, IBP2, and NIBP in 15-min Interval mode.
Battery charge time:
3.5 hours for 90% charge and 5.5 hours for full charge (equipment powered off)
5.5 hours for 90% charge and 8.5 hours for full charge (equipment powered on)
29.17.1 Recorder Performance Specifications
The recorder can print physiological waveforms.
Recorder paper: 50mm wide spooled thermal strip chart paper, minimum 200 ft long,
max diameter 50 mm when rolled.
Resolution: 600 dots per 25 mm at a print speed of 25mm/s, 400 dots at all other
print speeds
Maximum traces: 3 traces on thermal strip chart paper
Printing speed: 5 printing speeds: 3.125, 6.25, 12.5, 25, and 50mm/sec, each within
±
5%
Paper installation: Auto feed
Sensitivity:
± 5% on each trace
Performance: IEC60601-2-27:2011, and IEC60601-2-25:2011