CCO/Hemo Configuration Continuous Cardiac Output (CCO) & Continuous Hemodynamics (Optional)
1 - 4 Addendum to V Series Operating Instructions
1.2.2 Configuring CCO/Hemo Alarms 
For a full explanation about alarm responses and instructions for accessing and using these alarm 
dialogs, refer to the Alarms chapter in the V Series Operating Instructions.
1.2.2.1 Alarm Limits: CCO/Hemo Tab
TABLE 1-1 lists the alarm limit ranges for the CCO parameter, which is contained in the CCO tab and 
Hemo tab. 
TABLE 1-1 
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
CCO/CCO STAT
Unit of Measure: L/min
Resolution: 0.1 L/min
Invalid data: “- -”
High Limit Off, 2.
0 to 20.0  Of
f, 2.0 to 20.0  N/A
Low Limit Off, 1.0 to 19.0   Of
f, 1.0 to 19.0  N/A
CCI/CCI STAT
Unit of Measure: L/min/m
2
Resolution: 0.1 L/min/m
2
Invalid data: “- -”
High Limit Off, 2.0 to 20.0  Of
f, 2.0 to 20.0  N/A
Low Limit Off, 1.0 to 19.0   Of
f, 1.0 to 19.0  N/A
SVR
Unit of Measure: dyne-sec/cm
5
Resolution: 1 dyne-sec/cm
5
 
Invalid data: “- -”
High Limit Off, 500 to 5000  Off, 500 to 5000  N/A
Lo
w Limit Off, 0 to 2000   Of
f, 0 to 2000  N/A
SVRI
Unit of Measure: dyne-sec/cm
5
/m
2
Resolution: 1 dyne-sec/cm
5
/m
2
 
Invalid data: “- -”
High Limit Off, 1000 
to 
9950 
Of
f, 1000 to 9950  N/A
Low Limit Off, 0 to 4000   Of
f, 0 to 4000  N/A
T Blood
Unit of Measure: °F
Resolution: 0.1 °F 
Invalid data: “- -”
High Limit Off, 95.
0 to 
110.0 
Of
f, 95.0 to 110.0  N/A
Low Limit Off, 80.0 to 
100.0 
Of
f, 80.0 to 100.0  N/A
T Blood
Unit of Measure: °C
Resolution: 0.1°C 
Invalid data: “- -”
High Limit Off,
 35.0 to 43.0  Of
f, 35.0 to 43.0  N/A
Low Limit Off, 26.0 to 38.0  Of
f, 26.0 to 38.0  N/A
* The factory default configuratio
n settings are in bold.