Addendum to V Series Operating Instructions 1 - 5
Continuous Cardiac Output (CCO) & Continuous Hemodynamics (Optional) CCO/Hemo Configuration
SV/SV STAT
Unit of Measure: ml/beat
Resolution: 1 ml /beat
Invalid data: “- -”
High Limit Off, 20 
to 300  Off, 20 to 300  N/A
Low Limit Off, 0 to 140  Of
f, 0 to 140 N/A
SVI/SVI STAT
Unit of Measure: ml/beat/m
2
Resolution: 0.1 ml/beat/m
2
Invalid data: “- -”
High Limit Off, 20.0 to 
200.0 
Of
f, 20.0 to 200.0  N/A
Low Limit Off, 0.0 to 100.0  Of
f, 0.0 to 100.0  N/A
EDV/EDV STAT
Unit of Measure: ml
Resolution: 1 ml 
Invalid data: “- -”
High Limit Off, 40 
to 800  Of
f, 40 to 800  N/A
Low Limit Off, 0 to 400
 
Of
f, 0 to 400  N/A
EDVI/EDVI STAT
Unit of Measure: ml/m
2
Resolution: 0.1 ml/m
2
 
Invalid data: “- -”
High Limit Off, 20.0 to 
400.0 
Of
f, 20.0 to 400.0  N/A
Low Limit Off, 0.0 to 200.0  Of
f, 0.0 to 200.0  N/A
RVEF/RVEF STAT
Unit of Measure: %
Resolution: 1%
Invalid data: “- -”
High Limit Off, 20 
to 100  Of
f, 20 to 100  N/A
Low Limit Off, 0 to 60
 
Of
f, 0 to 60  N/A
TABLE 1-1 
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
* The factory default configuration settings are in bold.