Display Options Troubleshooting Display Setup
3 - 20 V Series Operating Instructions
3.3 Display Options Troubleshooting
MESSAGE/PROBLEM REASON SOLUTION
Cannot copy display preset. An Empty preset was selected. Select a configured preset. 
The Copy button is disabled 
when the preset is selected. 
An Empty preset was selected. Select a configured preset. 
There is no image on the display.  Monitor is off.  Turn monitor on.
Monitor is not properly seated in 
V Dock.
Reseat monitor in V Dock.
Monitor is not plugged into 
power source.
Plug monitor into power source. 
Desired parameters do not 
display on monitor.
Parameter(s) not included in 
Display Preset.
Refer to "Configuring Display 
Presets" on page 3-4 for 
instructions on modifying the 
preset.
Monitor is blank and reads 
“Standby mode.”
The Standby button was selected 
in the Display Options dialog.
Touch the display to resume 
monitoring. 
Touch screen is difficult to 
maneuver. 
Display needs to be calibrated.  Refer to "Configuring Display 
Presets" on page 3-4 for 
instructions on touch screen 
calibration.
Only see ECG leads in the 
waveform tiles. 
Display is in the View All ECG 
mode. 
Select the View All ECG button in 
the Display Options dialog. 
Overlap button in Display Setup 
dialog not enabled.
A digital parameter tile is selected. Select a waveform tile. 
A waveform tile is selected but it 
is not directly above or below 
another IBP waveform channel. 
Select a waveform tile that is 
directly above or below an existing 
IBP waveform channel.