Alarm Limits Alarms
8 - 6 V Series Operating Instructions
8.2.2.2 Pressure Tab
TABLE 8-3 lists the alarm limit ranges for the parameters contained in the Pressure tab.
Apnea Alarm NOTE: Refer to "Configuring Apnea Alarms" on page 10-5 for additional
information.
Apnea Latch
Apnea Delay
Unit of Measure: seconds
Resolution: 1second
N/A 10 to 60
Default: 30
10 to 20
Default: 15
10 to 20
Default: 15
TABLE 8-2
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
TABLE 8-3
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
NIBP
Unit of Measure: mmHg
Resolution: 1 mmHg
Invalid data: “- -
High Systolic
Limit
Off, 60 to 240
Default: Off
Off, 60 to 140
Default: Off
Off, 50 to 100
Default: Off
Low Systolic
Limit
Off, 55 to 150
Default: Off
Off, 55 to 120
Default: Off
Off, 45 to 90
Default: Off
High
Diastolic
Off, 40 to 150
Default: Off
Off, 40 to 120
Default: Off
Off, 40 to 90
Default: Off
Low
Diastolic
Off, 30 to 140
Default: Off
Off, 30 to 100
Default: Off
Off, 20 to 80
Default: Off
High
Mean
Off, 60 to 200
Default: Off
Off, 40 to 140
Default: Off
Off, 40 to 90
Default: Off
Low
Mean
Off, 39 to 150
Default: Off
Off, 39 to 100
Default: Off
Off,28 to 70
Default: Off
IBP 1-8
Unit of Measure: mmHg
Resolution: 1 mmHg
Invalid data: “- -
High Systolic
Limit
Off, 5 to 300
Default: Off
Off, 5 to 300
Default: Off
Off, 5 to 300
Default: Off
Low Systolic
Limit
Off, 0 to 150
Default: Off
Off, 0 to 150
Default: Off
Off, 0 to 150
Default: Off
High
Diastolic
Off, 5 to 200
Default: Off
Off, 5 to 200
Default: Off
Off, 5 to 200
Default: Off
Low
Diastolic
Off, 0 to 150
Default: Off
O
f
f, 0 to 150
Default: Off
Off, 0 to 150
Default: Off
High
Mean
Off, 5 to 200
Default: Off
Off, 5 to 200
Default: Off
Off, 5 to 200
Default: Off
Low
Mean
Off, 0 to 150
Default: Off
Off, 0 to 150
Default: Off
Off, 0 to 150
Default: Off

Table of Contents

Other manuals for Mindray V Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V Series and is the answer not in the manual?

Mindray V Series Specifications

General IconGeneral
BrandMindray
ModelV Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals