Alarm Limits Alarms
8 - 8 V Series Operating Instructions
CO2 ET
Unit of Measure: mmHg
Resolution: 1 mmHg
High Limit Off, 20 to 80
Default: 60
Off, 20 to 80
Default: 60
Off, 20 to 80
Default: 60
Low Limit Off, 5 to 50
Default:15
Off, 5 to 50
Default:15
Off, 5 to 50
Default:15
CO2 ET
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 2.0 to 10.0
Default:10.0
Off, 2.0 to 10.0
Default:10.0
Off, 2.0 to 10.0
Default:10.0
Low Limit Off, 1.0 to 6.0
Default: 2.0
Off, 1.0 to 6.0
Default: 2.0
Off, 1.0 to 6.0
Default: 2.0
CO2 ET
Unit of Measure: kPa
Resolution: 0.1 kPa
High Limit Off, 2.0 to 10.0
Default: 10.0
Off, 2.0 to 10.0
Default: 10.0
Off, 2.0 to 10.0
Default: 10.0
Low Limit Off, 1.0 to 6.0
Default: 2.0
Off, 1.0 to 6.0
Default: 2.0
Off, 1.0 to 6.0
Default: 2.0
CO2 INSP
Unit of Measure: mmHg
Resolution: 1 mmHg
High Limit Off, 5 to 30
Default: 10
Off, 5 to 30
Default: 10
Off, 5 to 30
Default: 10
Low Limit N/A N/A N/A
CO2 INSP
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 1.0 to 4.0
Default: 1.0
Off, 1.0 to 4.0
Default: 1.0
Off, 1.0 to 4.0
Default: 1.0
Low Limit N/A N/A N/A
CO2 INSP
Unit of Measure: kPa
Resolution: 0.1 kPa
High Limit Off, 1.0 to 4.0
Default: 1.0
Off, 1.0 to 4.0
Default: 1.0
Off, 1.0 to 4.0
Default: 1.0
Low Limit N/A N/A N/A
Des Insp
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 18.0
Default: 6.0
Off, 0.2 to 18.0
Default: 6.0
Off, 0.2 to 18.0
Default: 6.0
Low Limit Off, 0.0 to 17.8
Default: Off
Off, 0.0 to 17.8
Default: Off
Off, 0.0 to 17.8
Default: Off
Des ET
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 18.0
Default: Off
Off, 0.2 to 18.0
Default: Off
Off, 0.2 to 18.0
Default: Off
Low Limit Off, 0.0 to 17.8
Default: Off
Off, 0.0 to 17.8
Default: Off
Off, 0.0 to 17.8
Default: Off
Enf Insp, Hal Insp, Iso
In
sp
Un
it of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 5.0
Default: 2.0
Off, 0.2 to 5.0
Default: 2.0
Off, 0.2 to 5.0
Default: 2.0
Low Limit Off, 0.0 to 4.8
Default: Off
Off, 0.0 to 4.8
Default: Off
Off, 0.0 to 4.8
Default: Off
Enf ET, Hal ET, Iso ET
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 5.0
Default: Off
Off, 0.2 to 5.0
Default: Off
Off, 0.2 to 5.0
Default: Off
Low Limit Off, 0.0 to 4.8
Default: Off
Off, 0.0 to 4.8
Default: Off
Off, 0.0 to 4.8
Default: Off
Sev Insp
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 8.0
Default: 5.0
Off, 0.2 to 8.0
Default: 5.0
Off, 0.2 to 8.0
Default: 5.0
Low Limit Off, 0.0 to 7.8
Default: Off
Off, 0.0 to 7.8
Default: Off
Off, 0.0 to 7.8
Default: Off
TABLE 8-5
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE