Specifications Product Specifications
29 - 10 V Series Operating Instructions
29.1.9 CO
2
Module Safety Designations:
TABLE 29-9
Type of protection against electric shock: Class II
Degree of protection against electric shock: CO
2
- Type BF, Non-Defib Proof
Supply Connection: +12VDC, +5VDC
Mode of Operation: Continuous
Protection Against Hazards of Explosion: Not protected (Ordinary)
Protection Against Ingress of Liquids: Drip-proof (IPX1)
Degree of electrical connection between
equipment and patient:
Equipment designed for direct electrical connection
to the patient
Degree of Mobility: Portable
Maximum Size: 1.988"H x 6.2"W x 5.5"D
Maximum Weight: 1.12 lbs.
Data Sample Rate: 80 Hz
Operating Temperature: +5°C to +40°C
Operating Humidity: 15% to 95%, non-condensing
Operating Altitude: 1060hPa to 700hPa
Storage Temperature: -20°C to +60°C
Storage Humidity: 10% to 95%, non-condensing
Storage Altitude: 1060hPa to 570hPa
Controls/Indicators: 1- LED, Powered and Communicating
Interfaces: 1- CO
2
Exhaust
1- CO
2
1- Module Bay (Mx)