Specifications Product Specifications
29 - 12 V Series Operating Instructions
29.1.12 Recorder Safety Designations:
The recorder module function is in accordance with the applicable requirements of IEC 60601-2-27
and IEC 60601-2-25.
Maximum Size: 0.988“H x 6.2”W x 5.5”D
Maximum Weight: 0.75 lbs.
Operating Temperature: +5°C to +40°C
Operating Humidity: 15% to 95%, non-condensing
Operating Altitude: 1060hPa to 700hPa
Storage Temperature: -20°C to +60°C
Storage Humidity: 10% to 95%, non-condensing
Storage Altitude: 1060hPa to 570hPa
Controls/Indicators: 1- LED, Powered and Communicating
2- LED, Channel Indicators
Interfaces: 1- Temp
1- Module Bay (Mx)
TABLE 29-11 (Continued)
TABLE 29-12
Type of protection against electric shock: Class II
Degree of protection against electric shock: Recorder Module- Not protected
Mode of Operation: Continuous
Supply Connection: +12VDC, +5VDC from Module Bay and V Hub
Protection Against Hazards of Explosion: Not protected (Ordinary)
Protection Against Ingress of Liquids: Drip-proof (IPX1)
Degree of electrical connection between equipment
and patient:
None
Degree of Mobility: Portable
Maximum Size: 2.996“H x 6.2”W x 5.5”D
Maximum Weight: 1.50 lbs.
Operating Temperature: +5°C to +40°C
Operating Humidity: 10% to 95%, non-condensing
Operating Altitude: 1060hPa to 700hPa
Storage Temperature: -20°C to +60°C
Storage Humidity: 15% to 95%, non-condensing
Storage Altitude: 1060hPa to 570hPa