Non-Invasive Blood Pressure (NIBP) Performance Product Specifications
29 - 32 V Series Operating Instructions
6. NIBP Systolic Pressure Alarms
NIBP system is equipped with High and Low NIBP rate alarms. Alarms can be triggered when
measured rate is equal to or greater/lesser than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
7. NIBP Diastolic Pressure
8. NIBP Diastolic Pressure Alarms
NIBP system is equipped with High and Low NIBP rate alarms. Alarms can be triggered when
measured rate is equal to or greater/lesser than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
9. NIBP Mean Alarms
NIBP system is equipped with High and Low NIBP rate alarms. Alarms can be triggered when
measured rate is equal to or greater/lesser than the alarm setting.
High Alarm Limits:
Adult: 60 to 240 mmHg in steps of 1 mmHg, default: Off
Pediatric: 60 to 140 mmHg in steps of 1 mmHg, default: Off
Neonate: 50 to 100 mmHg in steps of 1 mmHg, default: Off
Adult: 55 to 150 mmHg in steps of 1 mmHg, default: Off
Pediatric: 55 to 120 mmHg in steps of 1 mmHg, default: Off
Neonate: 45 to 90 mmHg in steps of 1 mmHg, default: Off
Accuracy: Mean error ±5 mmHg
Standard Deviation: less than 8 mmHg
Measurement Range: Hg Adult Mode: 30 to 200 mmHg
Pediatric Mode: 30 to 150 mmHg
Neonate Mode: 20 to 100 mmHg
Resolution: 1 mmHg
Adult: 40 to 150 mmHg in steps of 1 mmHg, default: Off
Pediatric: 40 to 120 mmHg in steps of 1 mmHg, default: Off
Neonate: 40 to 90 mmHg in steps of 1 mmHg, default: Off
Adult: 30 to 140 mmHg in steps of 1 mmHg, default: Off
Pediatric: 30 to 100 mmHg in steps of 1 mmHg, default: Off
Neonate: 20 to 80 mmHg in steps of 1 mmHg, default: Off