Carbon Dioxide (CO2) Performance Product Specifications
29 - 44 V Series Operating Instructions
For all non-humidified FilterLine sets, Smart CapnoLine sets, and NIV-Line sets, the interval is up to
8 hours.
For all humidified FilterLine sets, Smart CapnoLine sets, and NIV-Line sets, the interval is greater
than 8 hours.
4. CO
2
Alarm Limits
The CO
2
module function includes High and Low CO
2
alarms. Alarms can be triggered when
measured CO
2
is equal to or greater/lesser than the alarm setting.
ETCO
2
High Alarm Limits:
ETCO
2
Low Alarm Limits:
Inspired CO
2
High Alarm Limits:
5. Respiration Alarm Limits
Refer to the alarms section ‘‘Impedance Respiration Performance’’ on page 29-28, which includes
apnea alarms.
Adult: 20 to 80 mmHg in steps of 1 mmHg, default: 60mmHg
2.0 to 10.0% in steps of 0.1%, default: 10%
2.0 to 10.0 kPa in steps of 0.1 kPa, default: 10 kpa
Pediatric: 20 to 80 mmHg in steps of 1 mmHg, default: 60 mmHg
2.0 to 10.0% in steps of 0.1%, default: 10%
2.0 to 10.0 kPa in steps of 0.1 kPa, default: 10 kPa
Neonate: 20 to 80 mmHg in steps of 1 mmHg, default: 60 mmHg
2.0 to 10.0% in steps of 0.1%, default: 10%
2.0 to 10.0 kPa in steps of 0.1 kPa, default: 10 kPa
Adult: 5 to 50 mmHg in steps of 1 mmHg, default: 15mmHg,
1.0 to 6.0% in steps of 0.1%, default: 2%,
1.0 to 6.0 kPa in steps of 0.1 kPa, default: 2 kPa
Pediatric: 5 to 50 mmHg in steps of 1 mmHg, default: 15 mmHg
1.0 to 6.0% in steps of 0.1%, default: 2%
1.0 to 6.0 kPa in steps of 0.1 kPa, default: 2 kPa
Neonate: 5 to 50 mmHg in steps of 1 mmHg, default: 15 mmHg
1.0 to 6.0% in steps of 0.1%, default: 2%
1.0 to 6.0 kPa in steps of 0.1 kPa, default: 2 kPa
Adult: 5 to 30 mmHg in steps of 1 mmHg, default: 10 mmHg
1.0 to 4.0% in steps of 0.1%, default: 1%
1.0 to 4.0 kPa in steps of 0.1 kPa, default: 0.1 kPa
Pediatric: 5 to 30 mmHg in steps of 1 mmHg, default: 10 mmHg
1.0 to 4.0% in steps of 0.1%, default: 1%
1.0 to 4.0 kPa in steps of 0.1 kPa, default: 0.1 kPa
Neonate: 5 to 30 mmHg in steps of 1 mmHg, default: 10 mmHg
1.0 to 4.0% in steps of 0.1%, default: 1%
1.0 to 4.0 kPa in steps of 0.1 kPa, default: 0.1 kPa