4.7.5.3.1 Set (dead load)
In this menu item (under Calib), the dead load/preload of the xBPI scale weighing
electronics is set.
Note:
For Minebea Intec both terms dead load and preload are used.
Requirements:
- The communication between the device and xBPI scale is active, and
- the scale parameters have been read, see Chapter 4.7.5.3.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [xBPI scale] - [Calib] - [Set] .
WP-C
R1 Max 29.9kg d= 0.0005kg
50000
–
kg
xBPI scale setup @admin
Reading parameters
read xBPI settings...
Ticks indicate the progress.
The following error message is displayed if communication with the xBPI scale is
not possible:
STOP: "Retrieve failed – timeout in communication"
The xBPI setup menu opens.
WP-C
R1 Max 29.9kg d= 0.0005kg
50000
+
kg
xBPI scale setup @admin
Dead load
Set
Delete
SPAN
Adjust with user weight
Adjust with automatic weight
4 System setup Process Controller Maxxis 5 PR 5900